Hioki AC/DC high voltage divider VT1005 | APC Technology Group

VT1005

VT1005

Hioki

AC/DC High Voltage Divider VT1005 - Safely measure high voltages up to 5000 V, detects efficiency improvement effects on the order of 0.1%

The Hioki VT1005 safely measures voltages up to 5000 V, converting them to a lower voltage then outputs them to a power analyser. As a result, the device can be used to measure the efficiency of inverters that convert high-voltage power. In addition, the Hioki VT1005 can measure the voltage across a band that ranges from DC to 4 MHz with an amplitude error of ±0.1% (from DC to 200 kHz) and phase error within ±0.1°(*1) (from DC to 500 kHz). The VT1005 measures not only the fundamental wave of the power output from the inverter but also the switching frequency component accurately and with high repeatability.
Contact us for lead time

Click below to request a quote for this item. Or for product advice, stock and lead time enquiries call our team on 0330 313 3220.

More Information
Attribute Value
Key Features
  • Noise Resistance: CMRR 80 dB Typical (100 kHz), Differential Input Method
  • Max. Input 5000 V (*1), 2000 V CAT II , 1500 V CAT III
  • Measurement Accuracy: ±0.08% (DC), ±0.04% (50/60 Hz), ±0.17% (50 kHz)
  • Frequency Flatness: ±0.1% Amplitude Band 200 kHz Typical
  • Frequency Flatness: ±0.1° Phase Band 500 kHz Typical (*1)
  • Measurement Band: DC to 4 MHz (-3 dB)
Basic Specifications
Maximum rated voltage 5000 V rms, ±7100 V peak (*1)
Maximum rated voltage
(line-to-ground)
No measurement category: 5000 V AC/DC (*2)
Measurement category II: 2000 V AC/DC (*3)
Measurement category III: 1500 V AC/DC (*4)
Voltage dividing ratio 1000:1
Measurement accuracy ±0.08%(DC), ±0.04%(50/60 Hz), ±0.17% (50 kHz)
Frequency
flatness
Band where amplitude falls within ±0.1% range: 200 kHz (typical)
Band where phase falls within ±0.1° range: 500 kHz (typical) (*5)
Measurement bandwidth DC to 4 MHz (Amplitude and phase accuracy specified up to 1 MHz)
Common-mode voltage
rejection ratio (CMRR)
50 Hz/60 Hz: 90 dB(Typical)
100 kHz: 80 dB (Typical)
Measurement method Differential input
Operating temperature and humidity range -10°C to 50°C (14°F to 122°F), 80% RH or less (non-condensing)
Storage temperature and humidity range −20°C to 60°C (-4°F to 140°F), 80% RH or less (non-condensing)
Power supply 100 V to 240 V AC (50/60 Hz)
Dimensions and Weight Approx. 195.0 mm (7.68 in) W × 83.2 mm (3.28 in) H × 346.0 mm (13.62 in) D mm
Approx. 2.2 kg (77.6 oz.)
Included accessories - L1050-01 Voltage Cord ×1 (1.6 m/ 5.25 ft)
- L9217 Connection Cord ×1 (insulated BNC, 1.6 m/ 5.25 ft)
- 9704 Conversion Adapter ×1 (insulated-female BNC-to-banana plug)
- Power cord ×1
Standards Safety: EN 61010
EMC: EN 61326 Class A
Instrument Type
More Information
Attribute Value
Data Sheet Click here to view

Related Products

  1. LR5051
  2. LR8513
  3. BT3564
  4. FT3701
  5. LR8514